Model Number :PS1DP-Cryo
Place of Origin :CHINA
Price :Price Negotiable | Contact us for a detailed quote
Payment Terms :T/T
Vacuum :Low-temperature Vacuum <1.2E-3Pa
Sample Temperature :<10 K-420 K
Vibration :Sample Stage Vibration <1 μm (peak-to-peak)
Needing Range :Any Point Within 25 Mm Diameter
Number of Probe Arms :Standard 4 Probe Arms, Up To 8 Supported
Probe Arm Stroke :X-Y-Z, 50 Mm-25 Mm-25 Mm
Temperature Stability :<±20 MK (10 K-420 K)
Magnetic Field :Superconducting Magnet, Vertical Direction, ±3 T Strength
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description

Cryogenic Vertical Superconducting Magnetic Field Probe Station

Product Introduction

The PS1DP-Cryo closed-loop cryogenic superconducting magnet probe station, based on a 4 K closed-loop cryogenic probe station, adds a ±3 T superconducting magnet, providing a low-temperature strong magnetic field environment for 2-inch samples. It can accommodate samples for electrical transport testing of 2D materials, spintronic devices, superconducting materials, and Hall effect testing of semiconductor materials or devices. The probe station provides comprehensive control and testing software for automated and systematic testing solutions. Users can monitor and control the temperature and magnetic field of the probe station in real-time via software, and perform testing sequences for magnetoresistance, second harmonic, Hall effect, SOT, etc., making it an efficient solution for electrical transport characterization under low-temperature strong magnetic fields.

Equipment Performance
Equipment Performance Indicator Description
Sample Temperature <10 K-420 K
Temperature Stability <±20 mK (10 K-420 K)
Magnetic Field Superconducting magnet, vertical direction, ±3 T strength
Vibration Sample stage vibration <1 μm (peak-to-peak)
Vacuum Low-temperature vacuum <1.2E-3Pa
Probe Arm Stroke X-Y-Z, 50 mm-25 mm-25 mm
Needing Range Any point within 25 mm diameter
Number of Probe Arms Standard 4 probe arms, up to 8 supported
Thermal Anchor Probe arms and sample stage temperature difference <10 K
Sample Space Sample holder diameter up to 51 mm
Sample Holder Type Ground, coaxial, triaxial options
DC Probe ZN50, tip material tungsten or beryllium copper
Frequency Range 0-50 MHz with low-temperature coaxial cable, 0-1 GHz with semi-rigid coaxial cable
Microwave Probe GSG, tip material tungsten
Frequency Range 0-40 GHz with K-type connector, 0-67 GHz with 1.85 mm connector
Electrical Leakage Current <100 fA@1 V
Optical Microscope 0.75 X-3.75 X continuous zoom